Chapter 4 Operation
51
10. Temperature ramp control ON/OFF
Select “OFF” and press the
ENT.
key. Then, press the
NEXT
key.
In ramp control, temperature rise or fall from the setting in the previous step to that of this step is
controlled at a constant rate, or ramp.
Temperature ramp control status
:
ON
Turns temperature ramp control ON.
OFF
Turns temperature ramp control OFF.
ENT.
:
Enters the selection.
Stands for “Temperature Ramp”.
○
PRGM SET
* Temperature ramp control cannot be used in temperature
pull-down with LH cabinets because the heater is shut OFF
in temperature pull-down. Temperature is pulled down
naturally.
* Temperature ramp control cannot be used when soak
control is used.
* Temperature ramp control cannot be used with step No. 1.
It is kept OFF even if turned ON in step No. 1 setup.
Press the
NEXT
key and move to the next setting.
11. Humidity ramp control ON/OFF (Not necessary for LU cabinet)
Select “OFF” and press the
ENT.
key. Then, press the
NEXT
key.
In ramp control, humidity rise or fall from the setting in the previous step to that of this step is
controlled at a constant rate, or ramp.
Humidity ramp control status
ON
Turns humidity ramp control ON.
OFF
Turns humidity ramp control OFF.
ENT.
Enters the selection.
Stands for “Humidity Ramp”.
PRGM SET
* Humidity ramp control cannot be used in humidity
pull-down with LH cabinets because the heater is shut OFF
in humidity pull-down. Humidity is pulled down naturally.
* Humidity ramp control cannot be used when soak control is
used.
* Humidity ramp control cannot be used with step No.1. It is
kept OFF even if turned ON in step No.1 Setup.
Press the
NEXT
key and move to the next setting.