Chapter 4 Operation
53
Step 3 Setup
2. Select the step edit content selection STEP/END/SAVE.
To setup the next step No., select “STEP” and press the
ENT.
key.
3. Input the setup No.
Input “3” and press the
ENT.
key. Then, press the
NEXT
key.
4. Turn temperature (humidity) link control ON.
Select “ON” and press the
ENT.
key. Then, press the
NEXT
key.
5. Set target temperature.
Input “85.0” and press the
ENT.
key. Then, press the
NEXT
key.
6. Turn humidity control ON.
Select “ON” and press the
ENT.
key. Then, press the
NEXT
key.
7. Set target humidity.
Input “50” and press the
ENT.
key. Then, press the
NEXT
key.
8. Set the step exposure time.
Input “1.30” and press the
ENT.
key. Then, press the
NEXT
key.
9. Turn soak control OFF.
Select “OFF” and press the
ENT.
key. Then, press the
NEXT
key.
10. Turn temperature ramp control OFF.
Select “OFF” and press the
ENT.
key. Then, press the
NEXT
key.
11. Turn humidity ramp control OFF.
Select “OFF” and press the
ENT.
key. Then, press the
NEXT
key.
12. Enter step setup.
Select “YES” and press the
ENT.
key. Then, press the
NEXT
key.
Step 4 Setup
2. Select the step edit content selection STEP/END/SAVE.
To setup the next step No., select “STEP” and press the
ENT.
key.
3. Input the setup No.
Input “4” and press the
ENT.
key. Then, press the
NEXT
key.
4. Turn temperature (humidity) link control ON.
Select “ON” and press the
ENT.
key. Then, press the
NEXT
key.
5. Set target temperature.
Input “20.0” and press the
ENT.
key. Then, press the
NEXT
key.
6. Turn humidity control OFF.
Select “OFF” and press the
ENT.
key. Then, press the
NEXT
key.