Chapter 4 Operation
33
•
After feeding the specimen cable through the cable port, cover the
port with the included rubber plug. Without the plug, outside air will
infiltrate the test area, which can keep temperature (humidity) from
reaching the target setting. Cut a notch the size of the cable in the
plug and fit the cable in the notch.
Specimen
Rubber plug
(Option)
Cable
To specimen power
supply control terminals
and power source
Fig. 4.7 Feeding specimen cables through the cable port
Use a specimen cable that can withstand test conditions. Refer to the
following table for possible cable types.
Table 4.1 Cable type and withstand temperature
Cable type
Withstand
temperature
Vinyl cable
55°C
Heat-proof vinyl cable
100°C
Styrene-butadiene cabtyre cable
70°C
Electron crosslinked heat-proof plastic cable
100°C
Glass-braid silicon shielded cable
200°C
•
When not using the cable port, cover it with the rubber plug from
the inside and the cap from the outside.
Cap
Fig. 4.8 When not using the cable port
Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com