Chapter 4 Operation
29
Chapter 4 Operation
This chapter explains test preparations and preliminary checks, test
startup and ending of the temperature (and humidity) cabinets.
Perform the work described in this chapter before every test.
4.1 Test Preparations
How to Arrange Specimens
The withstand load of each included shelf is 5 kg evenly
distributed. Remember to spread specimens out across the entire
shelf without crowding them in any one place.
Space specimens apart and away from walls so as not to block air
circulation inside the test area.
Temperature(& humidity) uniformity drops if air cannot flow freely,
which will throw off test results.
You may freely change the position of the shelves to suit specimen
size and quantity. Arrange specimens so as not to obstruct air
currents inside the test area.
Setting Shelves and Shelf Bracket
1. Attach the shelf brackets in the order of the numbers shown below.
Ensure sufficient clearance from the ceiling or other shelves.
Supports have hooks spaced every 50 mm apart.
Shelf
bracket
①
②
③
Fig. 4.1 How to install shelf bracket
Note
Procedure