Chapter 6 Checks and Maintenance
76
Chapter 6 Checks and
Maintenance
This chapter explains equipment checks and maintenance. To keep the
cabinet in good working condition, perform checks and maintenance
periodically.
6.1 Check and Maintenance Lists
Check list
For an explanation on each check item, see "6.2 Checks".
If any of the following checks result bad, contact the place of purchase or
ESPEC CORP.
Table 6.1 Check list
Check item
When to check
Main power switch
(leakage breaker) trip test
Once monthly
Overheat protector trip test
・
Before long test runs
・
Before unmanned tests
Humidifying tray and regulator
water level check (Not necessary
for LU cabinet)
・
Once every three months
・
When cabinet is relocated. See
“3.4 Water Level Check”.
Maintenance list
For an explanation on each maintenance item, see "6.3 Maintenance".
Table 6.2 Maintenance list
Maintenance item
When to perform
Condenser cleaning (Not
necessary for LH cabinet)
Once monthly
Water tank cleaning (Not
necessary for LU cabinet)
Once monthly
Electromagnetic pump (water
tank) protective strainer
cleaning (Not necessary for LU
cabinet)
Once monthly
Humidifying tray cleaning (Not
necessary for LU cabinet)
Once monthly
Test area cleaning
Before every test
Water circuit (heat exhaust)
compartment cleaning
Once yearly
Take-down
Before long periods of disuse