Chapter 5 Getting More Out of Your Cabinet
73
12. Wet-bulb temperature calibration (Not displayed with LU cabinet)
Stands for
Calibration Wet Temperature
.
Amount of offset
Change the value.
Selects the digit.
ENT.
Enters the selection.
* Setting range: -5.0
wet-bulb temperature offset
+5.0.
SYSTEM
•
How to read the current wet-bulb temperature
Press the
key until the far right digit starts flashing.
At that moment, the wet-bulb temperature will be
displayed in the message area for 1 to 2 seconds.
Wet-bulb
(test area)
temperature
○
SYSTEM
Calibration
temperature
Pressing the
NEXT
key will move you to the user settings top display.
1. User setup top display