Liquiline System CA80CR
Operation
Hauser
69
10.1.4 Logbooks
Logbooks record the following events:
• Calibration/adjustment events
• Operation events
• Diagnostic events
Here you define how the logbooks should store the data.
In addition, you can also define individual data logbooks for sensors. Assign the logbook name
and select the measured value to be recorded. You can configure the recording rate (Scan time)
individually for every data logbook.
Further information on the logbooks
Data logbooks only apply for sensors (optional).
There are special data logbooks for the analyzer. These logbooks are automatically enabled
and assigned to the measuring channel.
SP1: Measuring channel 1 of the analyzer is assigned to data logbook SP1.
Diagnostics menu
Options
• Disabled
• Enabled
Factory setting
Disabled
Determine if, when the menu is called up, a hold status
should be issued at the current output.
Calibration active
Options
• Disabled
• Enabled
Factory setting
Enabled
Determine if, when the menu is called up, a hold status
should be issued at the current output.
External hold
Options
• None
• EtherNet/IP DO 1 to 8
Factory setting
None
Path: Menu/Setup/General settings/Logbooks
Function
Options
Info
Logbook ident
Customized text, 16 charac-
ters
Part of the file name when exporting a logbook
Path: Menu/Setup/General settings/Hold settings
Function
Options
Info