Liquiline System CA80CR
Commissioning
Hauser
61
Current action for
Read only
Is only displayed if an action is currently running.
Display in MM:SS
Start measurement
Starts a measurement immediately.
Is only displayed if no measurement is currently run-
ning.
Stop measurement
Stops the measurement; any sample or reaction solution
is discarded and the system is rinsed with the sample.
Afterwards the analyzer is set to a defined state.
Is only displayed if a measurement is currently running.
Determine zero point
Starts zero point calibration immediately with the set
stability criteria. The values of the calibration can be
accepted for an adjustment.
Determine calibration factor
Starts a calibration immediately with the configured sta-
bility criteria and the specified concentration value for
the standard solution. The values of the calibration can
be accepted for an adjustment.
Stop calibration
Stops the calibration currently underway; any sample or
reaction solution is discarded and the system is rinsed
with the standard solution. The analyzer is set to a
defined state.
Sample preparation 1
(CAT820, CAT860)
In the case of two-channel devices, Sample preparation
2 is also displayed.
Start rinse reagents
Maintenance
Bottle change
Bottle removal
Bottle selection
Options
• None
• Cleaner
• Standard 1
• Reagent 1
• Reagent 2
Select the bottles that you would like to remove.
It is possible to select multiple bottles.
RB = R1 (reagent 1)
RK = R2 (reagent 2)
If at least one
Bottle selection
is selected:
Confirm bottles removed
Bottle insertion
Bottle selection
Options
• None
• Cleaner
• Standard 1
• Reagent 1
• Reagent 2
Select the bottles that you would like to insert.
It is possible to select multiple bottles.
RB = R1 (reagent 1)
RK = R2 (reagent 2)
If at least one
Bottle selection
is selected:
Path: Menu/Operation
Function
Options
Info