Liquiline System CA80CR
Operation
Hauser
123
The hardware required for "Chemoclean Plus", such as control valves, pumps, compressed
air supply, media etc., must be provided by the customer.
Defining the cleaning cycle
Path: Menu/Setup/Additional functions/Cleaning/Cleaning 1 ... 4
Function
Options
Info
Cleaning cycle
Options
• Off
• Interval
• Weekly program
Factory setting
Off
Choose from a cleaning routine that is restarted at set
intervals and a user-definable weekly program.
Cleaning cycle = "Interval"
Cleaning interval
00-00:01 to 07-00:00
(DD-hh:mm)
Factory setting
01-00:00
The interval value can be between 1 minute and 7 days.
Example: You have set the value "01-00:00". Each day,
the cleaning cycle starts at the same time you started the
first cleaning cycle.
Cleaning cycle = "Weekly program"
Daily event times
00:00 to 23:59
(HH:MM)
1. "Event time 1 to 6": Define up to 6 times.
You can then choose from these for each
weekday afterwards.
2. For each day of the week, individually choose which
of the 6 times is to be used for a cleaning routine on
this particular day.
In this way you create weekly programs that are perfec-
tly adapted to your process.
Weekdays
Monday ... Sunday
Содержание Liquiline System CA80CR
Страница 2: ......
Страница 188: ...www addresses endress com ...