Operation
Liquiline System CA80CR
70
Hauser
Event logbook
Options
• Off
• Ring buffer
• Fill up buffer
Factory setting
Ring buffer
All diagnostic messages are recorded
Ring buffer
If the memory is full, the latest entry automatically over-
writes the oldest entry.
Fill up buffer
If the memory is full, there is an overflow, i.e. no new
values can be saved. The controller displays a correspon-
ding diagnostic message. The memory then has to be
cleared manually.
Analyzer event logbook
Options
• Off
• Ring buffer
• Fill up buffer
Factory setting
Ring buffer
All program messages are recorded
Ring buffer
If the memory is full, the latest entry automatically over-
writes the oldest entry.
Fill up buffer
If the memory is full, there is an overflow, i.e. no new
values can be saved. The controller displays a correspon-
ding diagnostic message. The memory then has to be
cleared manually.
Overflow warnings
Event logbook="Fill up buffer"
Calibration logbook
Options
• Off
• On
Factory setting
Off
Decide whether you want to receive a diagnostic mes-
sage from the controller in the event of memory overrun
of the logbook in question.
Diagnostic logbook
Configuration logbook
Overflow warnings
Analyzer event logbook="Fill
up buffer"
Options
• Off
• On
Factory setting
Off
If the fill-up buffer overruns,
you can decide whether you
want to receive a diagnostic
message from the controller
or not for each individual log-
book.
Analyzer data logbooks
For analyzer measured data
The entry is made automatically on completion of the
measurement. A setting does not need to be made. The
logbook is activated automatically. SP1 is assigned to
data logbook SP1.
Data logbook SP1
Assignment to the measuring channel
Source of data
Read only
The assigned measuring channel is displayed
Path: Menu/Setup/General settings/Logbooks
Function
Options
Info