Liquiline System CA80CR
Operation
Hauser
127
Dual conductivity
You can subtract two conductivity values and use the result, for example, to monitor the
efficiency of an ion exchanger.
Path: Menu/Setup/Additional functions/Mathematical functions/MF 1 to 6/Mode/Degassed conductivity
Function
Options
Info
Calculation
Options
• Off
• On
Factory setting
Off
On/off switch for the function
Cation conductivity
Connected conductivity sen-
sor
"Cation conductivity" is the sensor downstream from the
cation exchanger and upstream from the "degassing
module", "Degassed conductivity" is the sensor at the out-
let of the degassing module.
Measured value interrogation is obsolete as you can only
choose conductivity.
Degassed conductivity
Connected conductivity sen-
sor
CO2 concentration
Read only
View this value in a user-defined measuring screen or
output the value via the current output.
Path: Menu/Setup/Additional functions/Mathematical functions/MF 1 to 6/Mode/Dual conductivity
Function
Options
Info
Calculation
Options
• Off
• On
Factory setting
Off
On/off switch for the function
Inlet
The options depend on the
sensors connected
Select the sensors that are to be used as the minuend
(Inlet, e.g. sensor upstream from the ion exchanger) and
subtrahend (Outlet, e.g. sensor downstream from the
ion exchanger).
Measured value
Outlet
Measured value
Main value format
Options
• Auto
• #
• #.#
• #.##
• #.###
Factory setting