Electrical connection
Liquiline System CA80CR
38
Hauser
6.5
Connecting the sensors
Only use terminated genuine cables.
a0003350
Fig. 38: Example of Memosens data cable CYK10
Connecting the ferrules of the sensor cable to the E basic module
The outer shield of the cable is grounded by means of the metal gland under the E basic module.
6.6
Optional sensor inputs, current outputs and relays
WARNING
!
Module not covered
No shock protection. Danger of electric shock!
‣
If you are modifying or extending your hardware, always fill the slots from left to right. Do
not leave any gaps.
‣
If not all the slots are occupied, always insert a dummy or end cover into the slot to the right
of the last module. This ensures the unit is shock-protected.
‣
Always ensure shock protection is guaranteed particularly in the case of relay modules (2R,
4R, AOR).
a0023172
Fig. 39: Terminal strip