Liquiline System CA80CR
Maintenance
Hauser
157
12.8 Decommissioning
If the analyzer has not been in operation for more than 5 days, it must be decommissioned to
avoid any damage to the device.
To do so, proceed as follows:
1. Open the analyzer.
2. Stop the sample supply.
3. Press the soft key "MODE" and select "Manual mode".
4. Wait until all the actions are finished and "Current mode - Manual" appears on the display.
5. Select "Menu/Operation/Maintenance/Decommissioning/Sample collector/Empty sample
collector/Start".
The system is rinsed automatically and drained. This operation takes about 4 minutes.
6. Open all the chemical bottles and carefully remove the hoses.
7. Dry the ends of the hoses with a clean paper towel.
8. Place the ends of the hoses into an empty beaker.
9. Once the operation has stopped, select
"Menu/Operation/Maintenance/Decommissioning/Rinse analyzer".
10. Immerse all the hoses (R1, R2, S1, C and P) in a beaker containing approx. 200 ml of distilled
water in order to rinse the system with water.
11. Once again select "Menu/Operation/Maintenance/Decommissioning/Rinse analyzer".
12. Remove the hoses from the beaker and dry them with a clean paper towel.
13. Select "MODE/Decommissioning/Rinse analyzer" once again to rinse the hoses with air.
14. The analyzer can now be disconnected from the mains supply.
Содержание Liquiline System CA80CR
Страница 2: ......
Страница 188: ...www addresses endress com ...