10.4 Current inputs
The input can be used as the data source for limit switches and logbooks, for example.
Furthermore, external values can be made available as set points for controllers.
Diag. code337
Read only
Warning limit
01-00 to 99-00 (DD-HH)
Factory setting
60-00
Reset settings
Resets all settings specific to sample preparation All
other settings are retained.
If Installation type
Pressurized pipe sampling
is selected (with Liquiline System CAT810, for example):
Sample preparation 1
In the case of a two-channel device, Sample preparation
2 is also displayed.
Cleaning valve
Options
• Not available
• Available
Factory setting
Not available
In the event of retrofitting, the valve can be activated
here.
If Cleaning valve
Available
is selected:
Filter backflush
Options
• On
• Off
Factory setting
On
If Filter backflush
On
is selected:
Cleaning interval
00:10 to 02:00 HH:MM
Factory setting
00:30
Compressed air or water cleaning (optional) for longer
filter maintenance intervals.
Cleaning duration
10 to 30 s
Factory setting
10 s
Cleaning duration of the automatic water or compressed
air cleaning system
Cleaning discard time
Factory setting
Discard time once cleaning is complete. If water is used
as the flushing medium, this must be replaced by a fresh
sample before starting the next measurement.
Reset settings
Resets all settings specific to sample preparation All
other settings are retained.
Path: Menu/Setup/Sample preparation
Function
Options
Info