Liquiline System CA80CR
Technical data
Hauser
175
15.2.2 Signal on alarm
Adjustable, as per NAMUR Recommendation NE 43
• In measuring range 0 to 20 mA:
Error current from 0 to 23 mA
• In measuring range 4 to 20 mA:
Error current from 2.4 to 23 mA
• Factory setting for error current for both measuring ranges:
21.5 mA
15.2.3 Load
max. 500
15.2.4 Transmission behavior
Linear
15.3 Current outputs, active
15.3.1 Span
0 to 23 mA
15.3.2 Signal characteristic
Linear
15.3.3 Electrical specification
Output voltage
Max. 24 V
Test voltage
500 V
15.3.4 Cable specification
Cable type
Recommended: shielded cable
Cable specification
Max. 2.5 mm
2
(14 AWG)