Liquiline System CA80CR
Operation
Hauser
121
Standard cleaning
Standard cleaning involves, for example, cleaning a sensor with compressed air, as in the case
of the ion-selective sensor CAS40D (connection of cleaning unit for CAS40D
41).
Chemoclean
An example of the use of the "Chemoclean" function is the use of the CYR10 injector unit to clean
pH glass sensors. (CYR10 connection
Chemoclean Plus
Use the "Chemoclean Plus" function for interval measurements in aggressive media or for regular
cleaning operations using acid or alkaline solutions or special cleaning agents.
Path: Menu/Setup/Additional functions/Cleaning/Cleaning 1 ... 4/Standard clean
Function
Options
Info
Cleaning time
5 to 600 s
Factory setting
10 s
Cleaning time ... Cleaning duration
The cleaning duration and interval depend on the pro-
cess and sensor. Determine the variables empirically or
based on experience.
Path: Menu/Setup/Additional functions/Cleaning/Cleaning 1 ... 4/Chemoclean
Function
Options
Info
Cleaning time
0 to 900 s
Factory setting
5 s
Cleaning duration
The cleaning duration, pre-rinse and post-rinse times
and the interval depend on the process and sensor.
Determine the variables empirically or based on experi-
ence.
Prerinse time
0 to 900 s
Factory setting
0 s
Postrinse time
0 to 900 s
Factory setting
0 s