Technical data
Liquiline System CA80CR
174
Hauser
Cable length
Max. 100 m (330 ft)
15.2 Output
15.2.1 Output signal
Depending on version:
• 2 x 0/4 to 20 mA, active, potentially isolated
• 4 x 0/4 to 20 mA, active, potentially isolated
• 6 x 0/4 to 20 mA, active, potentially isolated
Modbus RS485
Ethernet and Modbus TCP
EtherNet/IP
Signal coding
EIA/TIA-485
Data transmission rate
2400, 4800, 9600, 19200, 38400, 57600 and 115200 Baud
Galvanic isolation
Yes
Bus termination
Internal slide switch with LED display
Signal coding
IEEE 802.3 (ethernet)
Data transmission rate
10 / 100 MBd
Galvanic isolation
Yes
IP address
DHCP or configuration via menu
Signal coding
IEEE 802.3 (ethernet)
Data transmission rate
10 / 100 MBd
Galvanic isolation
Yes
Connection
RJ45, optional M12 (D-encoded)
IP address
DHCP (default) or configuration via menu