Liquiline System CA80CR
Operation
Hauser
99
Path: Menu/Setup/Outputs/Current output x:y (slot:output number)
Function
Options
Info
Current output
Options
• On
• Off
Factory setting
Off
Use this function to activate and deactivate a variable
being output at the current output
Source of data
Options
• None
• Connected inputs
• Controller
Factory setting
None
The sources of data on offer depend on your device ver-
sion. The main value of the analyzer and all the sensors
and controllers connected to inputs are available for
selection.
Measured value
Options
• None
• Depends on theSource of
data
Factory setting
None
The measured value you can select depends on the
option selected under "Source of data".
The table on the next page lists the measured values available depending on the data source.
Apart from measured values from connected sensors, you can also select a controller as the data source. The best
way to do so is by using the "Additional functions" menu. Here, you can also select and configure the current out-
put for outputting the controlled variable.
Range lower value
Selection range and factory
settings depend on:
"Measured value"
You can output the entire measuring range of the "Mea-
sured value" or just some of it at the current output. For
this purpose, specify the start and end of the range in
accordance with your requirements.
Range upper value
Hold behavior (for sensors)
Options
• Freeze
• Fixed value
• None
Factory setting
Depends on the chan-
nel:output
Freeze
The device freezes the last current value.
Fixed value
You define a fixed current value that is output at the out-
put.
None
A hold does not affect this current output.
Hold current (for sensors)
Hold behavior="Fixed value"
0.0 to 23.0 mA
Factory setting
22.0 mA
Specify which current should be output at this current
output in the event of a hold.
Measured valuedepending on theSource of data
Source of data
Measured value
Options
• Main value
Содержание Liquiline System CA80CR