Liquiline System CA80CR
Electrical connection
Hauser
29
6.1.2
Remote operation via Ethernet/Web server/Modbus TCP/EtherNet/IP
6.2
Connecting the analyzer
NOTICE
The device does not have a mains switch.
‣
You must install the device near (distance < 3 m (10 ft)) an easily accessible and fused plug
socket, so that the device can be disconnected from the supply network.
Depending on the order code, the mounted cable glands with an M-thread should be
replaced by the G' or NPT glands enclosed.
This does not affect the M32 hose glands.
The specifications for protective grounding must be observed during installation.
6.2.1
Routing the cables
• Remove the bottle tray by lifting up the recessed grip slightly and then pulling it towards the
front.
• Remove the cover that is hooked into place.
• Using an Allen key, release the six screws on the carrier board and fold it out towards the
front. For convenience of handling, hook the carrier board onto the locking plate.
• Using a Phillips head screwdriver, release the six screws on the electronics compartment cover
and fold out the cover towards the front.
• Cable glands are available for the cable entry.
a0015876
Fig. 22: Modbus TCP and / or ethernet