Liquiline System CA80CR
Operation
Hauser
87
Cleaning start condition
(CAT860)
Options
• Immediate
• Date/time
• Disabled
Immediate:
Cleaning starts immediately.
Date/time:
Cleaning starts at the selected date/time.
Disabled:
Cleaning is not active.
When Start condition
Immediate
or
Date/time
is selected:
Cleaning interval
(CAT860)
00-01 to 90-00 (DD-HH)
Factory setting
01-00
Compressed air or water cleaning for longer filter main-
tenance intervals
Cleaning with liquid cleaner
If Start condition
Date/time
is selected:
Date
(CAT860)
01.01.1970 to 07.02.2106
DD.MM.YYYY
Factory setting
01.01.1970
Time
(CAT860)
00:00:00 to 23:59:59
HH:MM:SS
Factory setting
00:00:00
Residence time
(CAT860)
00:30 to 20:00 (MM:SS)
Factory setting
01:00
Reaction time of cleaner during the cleaning
Sampling pump interval
(CAT820, CAT860)
Pump time
Factory setting
10 s
Ratio between the pumping duration and the pumping
interval of the peristaltic pump. This affects the volume
of the pumped sample.
Pump pause
Factory setting
30 s
Pressurized air cleaning
(CAT820, CAT860)
Compressed air
(CAT820)
Options
• Available
• Not available
Factory setting
Depends on the device ver-
sion
Settings for sample preparation with or without com-
pressed air cleaning.
In the event of retrofitting, the compressed air can be
activated here.
Cleaning mode
Options
• On
• Off
Factory setting
On
Activates or deactivates automatic compressed air clea-
ning of the pump-to-filter hose and of the filter.
Path: Menu/Setup/Sample preparation
Function
Options
Info