Operation
Liquiline System CA80CR
76
Hauser
Modbus
EtherNet/IP
EtherNet Industrial Protocol is an open fieldbus standard of the Open DeviceNet Vendor
Association (ODVA).
→
www.odva.org
Path: Menu/Setup/General settings/Extended setup/Modbus
Function
Options
Info
Enable
Options
• Off
• On
Factory setting
Off
You can switch off Modbus communication at this point.
The software can then only be accessed via local opera-
tion.
Termination
Read only
If the device is the last in the bus, you can terminate via
the hardware.
--> "Electrical connection" section
Settings
Transmission Mode
Options
• TCP
• RTU
• ASCII
The transmission mode is displayed depending on the
version ordered. In the case of RS485 transmission, you
can choose between "RTU" and "ASCII".
Watchdog
0 to 999 s
Factory setting
5 s
If no data transmission takes place for longer than the
time set, this is an indicator that communication has
been interrupted.
After this time, input values received via the Modbus are
considered to be invalid.
Path: Menu/Setup/General settings/Extended setup/Ethernet
Function
Options
Info
Enable
Options
• Off
• On
Factory setting
On
You can switch ethernet communication on and off at
this point.
Settings
Link settings
Options
• Auto negotiation
• 10MBps Half duplex
• 10MBps Full duplex
• 100MBps Half duplex
• 100MBps Full duplex
Factory setting
Auto negotiation
Transmission methods of the communication channels:
• Full duplex:
simultaneous transmission of data in both
directions (simultaneous sending and receiving)
• Half-duplex:
alternating transmission of data in both directions
(alternating sending
or receiving)