Maintenance
Liquiline System CA80CR
156
Hauser
17. bottle tray back into the housing.
18. Connect the bottles to the appropriate covers and hoses.
19. Switch the analyzer back on.
20. Select "Menu/Operation/Maintenance/Bottle replacement/Bottle insertion/Bottle
selection".
21. Select all the bottles and press "OK" to confirm.
22. Then select "Confirm bottles inserted".
23. The system must be calibrated after a replacement routine. Select "Menu/Setup/Manual
operation/Determine calibration factor".
24. Press the soft key "MODE" and select "Continue automatic mode" or "MODE/Start automatic
mode".