151
12.1.4 Sample collector (optional)
Cleaning interval:
• Depends on the application; recommendation: every 3 months
1. Open the analyzer.
2. Press the soft key "MODE" and select "Manual mode".
3. Wait until all operations have stopped.
4. Disconnect or stop pumping of the sample.
5. Select "Menu/Operation/Maintenance/Decommissioning/Sample collector/Empty sample
collector" and press "Start".
There should not be any more sample left in the sample collecting vessel.
6. Remove all the hoses from the sample collecting vessel and the cable of the level monitoring
system and take the sample collecting vessel out of the holder unit.
7. Open the sample collector by turning the cover in the clockwise direction.
8. Clean it with a small brush and plenty of water.
9. Return the collector to the holder unit and connect up all the hoses again correctly.
10. Start the sample preparation system.
11. Press the soft key MODE and select "Continue automatic mode" to start normal
measurement.
Replace the O-ring in the cover of the sample collection vessel annually. Grease the O-ring
before inserting it. The O-ring and silicone grease are included in the maintenance kit.
12.2 Replacing reagents, standard and cleaning agent
1. Press the soft key "MODE" and select "Manual mode".
2. Wait until all operations have stopped.
3. Remove the bottles that you wish to replace by removing the hoses at the liquid manager.
Use a paper towel to catch any chemicals that may escape and replace them with fresh
reagent, standard solution or cleaning agent.
4. Select "Menu/Operation/Maintenance/Bottle change mode/Bottle removal/Bottle
selection".
5. Select the bottles that you want to remove and click OK to confirm.
Replacement interval for CA80CR:
Reagent CY80CR
every 3.5 months (typically, measuring interval 10 min,
depending on temperature and measuring range)
Standard solution CY80CR• every 2.4 months (when cleaning interval is 48 h)
Cleaner CY800
every 4.5 months (when cleaning interval is 48 h)