Operation
Liquiline System CA80CR
88
Hauser
Cleaning interval
00:30 to 04:00 HH:MM
Factory setting
02:00
Cleaning interval of the automatic compressed air clea-
ning system
Cleaning duration
10 to 60 s
Factory setting
30 s
Cleaning duration of the automatic compressed air clea-
ning system
Heatings
(CAT820, CAT860)
Cabinet
Options
• Available
• Not available
Factory setting
Depends on the device ver-
sion
In the event of retrofitting, the heating can be activated
here.
Hose filter
Options
• Available
• Not available
Factory setting
Depends on the device ver-
sion
In the event of retrofitting, the heating can be activated
here.
Hose analyzer
Options
• Available
• Not available
Factory setting
Depends on the device ver-
sion
In the event of retrofitting, the heating can be activated
here.
Diagnostics settings
Limit change filter
(CAT820, CAT860)
Function
Options
• Off
• On
Factory setting
Off
Diag. code729
Read only
Warning limit
01-00 to 99-00 (DD-HH)
Factory setting
60-00
Limit change hose
(CAT820, CAT860)
Function
Options
• Off
• On
Factory setting
On
Path: Menu/Setup/Sample preparation
Function
Options
Info