Liquiline System CA80CR
Operation
Hauser
105
Outputting the status of a cleaning function
Function="Diagnostics"
Function
Options
Info
Operating mode
Options
• as assigned
• Namur M
• Namur S
• Namur C
• Namur F
Factory setting
as assigned
as assigned
If this option is selected, the diagnostics messages which
you have individually assigned to the relay are output via
the relay.
Namur M ... F
If you decided to use one of the Namur classes, all the
messages that are assigned to the individual class are
output via the relay.
You can also change the Namur class assignment for
every diagnostics message.
(Menu/Setup/General settings/Diagnostics/Device
behavior or Menu/Setup/Inputs/../Diagnostics set-
tings/Diag. behavior)
Attributed diagnostic messa-
ges
Operating mode="as assigned"
Read only
List of diagnostic messages
All the messages assigned to the relay output appear on
the display.
You do not have the option of editing the information
here.
Function="Cleaning" (for sensors)
Function
Options
Info
Assignment
Options
• None
• Cleaning 1 - Water
• Cleaning 1 - Cleaner
• Cleaning 2 - Water
• Cleaning 2 - Cleaner
• Cleaning 3 - Water
• Cleaning 3 - Cleaner
• Cleaning 4 - Water
• Cleaning 4 - Cleaner
Factory setting
None
Here you can specify how a cleaning function should be
displayed for the alarm relay.
You can define the cleaning programs under:
Menu/Setup/Additional functions/Cleaning.