Electrical connection
Liquiline System CA80CR
32
Hauser
6.2.4
Cable terminals
Plug-in terminals for Memosens and RS485 connections
After connection, make sure that every cable end is securely in place. Terminated cable
ends, in particular, tend to come loose easily if they have not been correctly inserted as far
as possible.
All other plug-in terminals
6.3
Connecting sample preparation
6.3.1
Connecting the optional valve of Liquiline System CAT810
1. Disconnect the mains plug.
2. Remove the bottle tray by lifting up the recessed grip slightly and then pulling it towards the
front.
3. Guide the cable through the cable gland.
a0012691
Fig. 24: Press the screwdriver against
the clip (opens the terminal)
a0012692
Fig. 25: Insert the cable until the limit
stop
a0012693
Fig. 26: Remove the screwdriver (closes
the terminal)
a0012694
Fig. 27: Insert the screwdriver (opens
the terminal)
a0012695
Fig. 28: Insert the cable until the limit
stop
a0012696
Fig. 29: Remove the screwdriver (closes
the terminal)