Liquiline System CA80CR
Diagnostics and troubleshooting
Hauser
135
302 Battery low
M
On
Off
Buffer battery of real time clock is low
It is located on the backplane. The date and time
are lost if the power is interrupted.
• Replace battery
304 Module data
F
On
On
At least 1 module has incomplete configuration
data
• Check module information
305 Power consumption
F
On
On
Total power consumption is too high
1. Check installation
2. Remove sensors/modules
3. Restart the device
306 Software error
F
On
On
Internal firmware error
Device configuration does not match firmware
• Contact the Service Department
360 Cooling/Heating
S
On
Off
Temperature range in the housing exceeded
1. Check installation conditions and the ambi-
ent temperature
2. Replace the FMAB1 module
361 Cooling/Heating
F
On
On
Cooling/heating module defective
Defined temperature range not reached. This
could affect the functionality of the reagents.
1. Check whether the insulation cap is seated
correctly over the reagents
2. Replace cooling/heating module
362 Photometer temp.
F
On
Off
Temperature of the photometer too high
1. Replace photometer
363 Photometer temp.
F
On
Off
Temperature of the photometer too low
1. Replace photometer
364 Timeout dosing
F
On
On
The liquid manager/dosing dispenser has timed
out.
Possible reasons:
• Photoelectric barrier defective
• Blocking
-> Contact the Service Department
370 Internal voltage
F
On
On
Internal 24-V voltage outside the valid range
--> Check supply voltage
373 Electronictemp. high
M
On
Off
Temperature of backplane electronics is high
--> Check ambient temperature and energy
consumption
374 Sensor check
F
On
Off
Sensor measuring sequence counter
No measurement signal from sensor
1. Check sensor, replace if necessary
2. Check sensor connection
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current