Liquiline System CA80CR
Electrical connection
Hauser
47
6.9
Guaranteeing the degree of protection
Only the mechanical and electrical connections that are described in this manual, and are
necessary for the required, designated application, may be established on the device supplied.
‣
Please play close attention when performing the work as degrees of protection individually
confirmed for this product (ingress protection (IP), electrical safety, EMC interference
immunity, explosion protection) can no longer be guaranteed in the case of the following,
for example:
– Leaving off covers
– Loose or insufficiently tightened cables/cable ends
individual types of protection permitted for this product (impermeability (IP), electrical
safety, EMC interference immunity) can no longer be guaranteed here.
6.10 Post-connection check
WARNING
!
Wiring errors
Incorrect wiring puts the safety of people and the measuring point at risk. The manufacturer
does not accept any responsibility for errors that result from failure to comply with the
instructions in this manual.
‣
Put the analyzer into operation only if you can answer
yes
to
all
of the following questions.
Device state and specifications
1.
Are the analyzer and cables free from damage on the outside?
2.
Are the hoses free from damage on the outside?
Electrical connection
3.
Are the mounted cables strain relieved?
Visual inspection of the liquid-bearing lines
4.
Are all the hose connections leak-tight?
5.
is the hose in the hose gland strain-relieved?
6.
Have the bottles with reagent and standard been inserted and connected?
7.
Inspect the hose connections. Use the hose connection diagram as a guide.