Operation
Liquiline System CA80CR
126
Hauser
rH value
To calculate the rH value, a pH sensor and an ORP sensor must be connected. It is irrelevant
whether you are using a pH glass sensor, an ISFET sensor or the pH electrode of an ISE sensor.
Degassed conductivity
Carbon dioxide from the air can be a contributing factor to the conductivity of a medium. The
degassed conductivity is the conductivity of the medium excluding the conductivity caused by
carbon dioxide.
In the power station industry, for example, it is advantageous to use the degassed conductivity:
• The percentage of conductivity caused by corrosion products or fouling in the feed water can
be determined as early as when the turbines are started. The system excludes the initially high
conductivity values resulting from the ingress of air.
• If carbon dioxide is regarded as non-corrosive, the live steam can be directed to the turbine far
earlier during startup.
• If the conductivity value increases during normal operation, it is possible to immediately
determine an ingress of coolant or air by calculating the degassed conductivity.
Deviation control
Options
• Off
• On
Factory setting
Off
You can monitor the redundancy. Specify an absolute
limit value that may not be exceeded.
Deviation limit
Depends on the selected
measured value
Redundancy
Read only
View this value in a user-defined measuring screen or
output the value via the current output.
Pfad: Menu/Setup/Additional functions/Mathematical functions/MF 1 to 6/Mode/rH calculation
Function
Options
Info
Calculation
Options
• Off
• On
Factory setting
Off
On/off switch for the function
pH source
Connected pH sensor
Set the input for the pH sensor and the input for the ORP
sensor. Measured value interrogation is obsolete as you
must select pH or ORP mV.
ORP source
Connected ORP sensor
Calculated rH
Read only
View this value in a user-defined measuring screen or
output the value via the current output.
Path: Menu/Setup/Additional functions/Mathematical functions/1 to 6/Mode/Redundancy
Function
Options
Info