Transport and storage
multi N/C pharma HT
116
}
Remove the reagent bottle, the drip tray and other loose accessories. Wipe off the
hose(s) with a clean paper towel.
CAUTION! The hoses contain acid residue.
}
Detach the canulas from the hoses. Put the canulas in the canula packaging.
NOTICE! Package the canulas with care. The canulas may bend.
}
Remove the hoses from the connections on the halogen trap. Remove the halogen
trap from the clamps.
}
Remove and empty the TIC condensate container.
}
Pack open hose ends in protective bags and secure them in the analyzer, for example
with adhesive tape.
Figure 65Components secured behind the front doors for transport
}
Open the left side wall:
– Unscrew the four attachment screws. The screws are captive and remain attached
to the wall.
– Remove the protective grounding. Set the side wall aside safely.
}
Carefully remove the condensation coil from the holder, empty it and set it aside
safely.
}
Remove the combustion tube.
}
Remove the combustion furnace.
}
Pack open hose ends inside the device in protective bags and secure them on the an-
alyzer with adhesive tape.
}
Close the left side wall of the analyzer:
– Attach the protective grounding to the side wall.
– First screw in the screws on the bottom side and then on the top side. Tighten the
screws in turns.
}
Position the top furnace cover and secure it with adhesive tape.
}
Close the front doors of the analyzer.
}
Carefully package the accessories. Ensure that the glass components are packed to
prevent breakage.
}
Package the analyzer and the accessories in the original packaging.
ü
The analyzer is securely packed for transport.