multi N/C pharma HT
Security
11
2.4
Safety instructions: during operation
2.4.1
Summary of safety instructions
The operator must make sure that the device and its safety equipment is in sound
condition each time before starting up the device. This applies in particular after
each modification or extension of the device or its repair.
Observe the following:
¡
The device may only be operated if all items of protective equipment (e.g. covers
in front of electronic components) are in place, properly installed and fully oper-
ational.
¡
The sound condition of the protection and safety equipment must be checked
regularly. Any defects must be corrected as soon as they occur.
¡
Protective and safety equipment must never be removed, modified or switched
off during operation.
¡
Always ensure free access to the main switch and to the emergency shutdown
switches and locks during operation.
¡
The ventilation equipment on the device must be in good working condition.
Covered ventilation grilles or slots etc. may cause the device to break down or
may cause damage to it.
¡
Modifications, conversions and extensions to the device are only permitted after
consultation with Analytik Jena. Unauthorized modifications can jeopardize the
device’s operational safety and may lead to limitations regarding the warranty
and access to customer service.
¡
Keep all combustible materials away from the device.
¡
The furnace operates at temperatures of 700 to 950 °C. Do not touch the hot
components (furnace, condensation coil) during or directly after operation.
¡
Caution when handling glass components. Risk of broken glass and therefore risk
of injury!
¡
Ensure that no liquid enters the interior of the device, for example at cable con-
nections. There is a danger of electric shock.
¡
There is a risk of injury in the movement range of the autosampler. For example,
hands or fingers may be crushed. Maintain a safety distance from the autosam-
pler during operation.
¡
The optional chemiluminescence detector (CLD) contains an ozone generator
which produces ozone (O
3
). When used in accordance with the intended use, the
downstream ozone decomposer destroys the toxic gas. Various safety measures
result in the automatic shut-down of the ozone generator. However, the follow-
ing still applies: if you notice a strong smell of ozone, switch the device off imme-
diately and inform customer service. In order to guarantee perfect and safe oper-
ation, Analytik Jena recommends an annual inspection and maintenance of the
detector by customer service.
2.4.2
Safety instructions – protection against explosion and fire
The device may not be operated in an explosive environment.
Smoking or handling open flames are prohibited in the room in which the device is
operated!