Installation and commissioning
multi N/C pharma HT
56
Figure 40Transport lock
1 Autosampler arm
2 Transport retaining clip
3 Screws
}
Install the stirring arm:
– Install the stirring arm on the bracket on the rear of the autosampler arm.
}
Screw on the arm with the supplied countersunk screws (M4x10) using the hexagon
head wrench (A/F2.5).
– Tighten the screws evenly to aligned the arm.
– Connect the stirrer cable to the "Stirrer" connection on the rear of the autosam-
pler.
Figure 41Installing the stirring arm
1 Bracket on the autosampler arm
2 Countersunk screws
3 Stirring arm
}
Place the autosampler next to the analyzer. Position the autosampler so that enough
space is provided behind the device for the motion range of the autosampler arm as
well.
}
Connect the low voltage side cable of the table power supply unit to the rear of the
autosampler. Do not connect the power supply unit to the mains power supply yet.
}
Connect the supplied serial data cable to the "sampler" interface on the rear of the
analyzer. Connect the other end of the data cable to the interface on the autosam-
pler.
}
Plug the grounding conductor into the connection on the rear of the analyzer.
}
Connect the waste hose to the wash cup of the autosampler and to a suitable waste
container or drain.
NOTICE! Route the outlet hose at a constant downward incline. If necessary,
shorten the hose. The hose must not dip into the liquid.
}
Install the wash cup on the autosampler.
}
Place the sample tray onto the space provided.