Troubleshooting
multi N/C pharma HT
114
¡
Faulty dosing.
¡
For manual sample supply: Ensure even
injection.
Error
Incorrect TNb analyses with
CLD (TC analy-
ses OK)
Cause
Remedy
¡
Faulty hose connection between the an-
alyzer and the detector.
¡
Ozone generator faulty.
¡
Check the hose connections.
¡
Inform the service.
Error
Condensate pump or phosphoric acid pump
leaky
Cause
Remedy
¡
Leaking hose connections.
¡
Defective pump hose.
¡
Check the connections.
¡
Replace the hose.
Error
5 V, 24 V indicator lamps on LED strip not
lighting
Cause
Remedy
¡
Power supply or electronics fault.
¡
Check the electrical connections.
¡
Check the laboratory power supply.
¡
Device fuse faulty.
¡
Inform the service.
Error
Status LED on the analyzer not lighting
(Lockin)
Cause
Remedy
¡
The internal program has not been
started.
¡
Switch the analyzer off and on again.
Error
Heating monitoring lamp on LED strip not
lighting
Cause
Remedy
¡
Incorrect temperature settings in the
software.
¡
Check the temperature settings under
Configuration |Options in the Analyzer
components tab.
¡
Faulty thermocouple (furnace). The
"Broken Thermocouple" lamp is lit on the
LED strip.
¡
Inform the service.
¡
Faulty electronics component.
¡
Inform the service.
¡
The combustion furnace is not con-
nected correctly.
¡
Check the connection of the combustion
furnace.