Installation and commissioning
57
}
Note the positioning of the tray. The label must be legible when facing the front of
the device. The two black centering pins on the contact surface of the autosampler
protrude into the drill holes on the bottom of the tray.
}
Insert the piercing canulas and holding-down clamps into the autosampler arm.
}
Clamp the two canulas high enough in the holder to prevent them dipping into the
vessels (basic position).
Figure 42Canula position for NPOC measurement with parallel (left) and non-paral-
lel (right) purging.
}
Secure the canulas by slightly tightening the knurled head screws.
}
Connect the hoses from the analyzer to the canulas using Fingertight connections:
– Hose AA – sample intake hose
Hose 7 – purging hose for NPOC measurement
– To do this, guide the hose through the banjo bolt (see image).
– Slide the conical nipple onto the hose with the conical side towards the banjo
bolt. The conical nipple and hose must be flush.
– Retighten the Fingertight connections.
1
2
3
Figure 43Fingertight connection
1 Hose
2 Banjo bolt
3 Conical nipple