Installation and commissioning
multi N/C pharma HT
36
4
Installation and commissioning
4.1
Installation conditions
4.1.1
Ambient conditions
¡
This laboratory device is designed for inside use.
¡
Avoid direct sunlight and radiation from heaters onto the device. If necessary, pro-
vide air conditioning.
¡
The installation site must be free of drafts, dust and caustic fumes.
¡
The room air must be as low in TOC and NO
x
as possible.
¡
Avoid mechanical shocks and vibrations.
¡
Do not locate the device near sources of electromagnetic interference.
¡
Place the device on a heat-resistant and acid-resistant surface.
¡
The device must be positioned in such a way that allows easy access from all sides.
¡
Keep the ventilation slits free and do not obstruct them with other devices.
The following climate requirements apply in the room of operation:
Operating temperature
+10 to 35 °C (air-conditioning recom-
mended)
Maximum humidity
90 % at 30 °C
Air pressure
0.7 to 1.06 bar
Storage temperature
5 to 55 °C
Humidity during storage
10 to 30 % (use desiccant)
Operating altitude (max.)
2000 m
4.1.2
Device layout and space requirements
The basic device has been designed as a compact bench-top device. The required space
depends on all components that make up the measuring station.
Further components of the measuring station:
¡
The PC, monitor and printer may be placed on a separate side table.
¡
An acid-resistant waste container can be placed on or under the bench.
Optional modules:
¡
The AS 10e and AS 21hp autosamplers are attached to the right side wall of the ba-
sic device. Alternatively, the autosamplers can be placed next to the analyzer on the
left.
¡
The AS vario and EPA Sampler autosamplers must be positioned to the right of the
device.
¡
The Swab-Test Module is attached to the left side wall of the basic device.
¡
The CLD nitrogen detector must be positioned to the left of the basic device.
Component
Dimensions (width x height x
depth)
Weight
Basic device
513 x 464 x 550 mm
30 kg
AS 10e autosampler
260 x 390 x 320 mm
4.5 kg
AS 21hp autosampler
260 x 390 x 320 mm
4.5 kg
AS vario autosampler
350 x 470 x 400 mm
15 kg
EPA Sampler
500 x 560 x 400 mm
15 kg