Troubleshooting
multi N/C pharma HT
110
Error code: Error message
Peltier temperature outside range
Cause
Remedy
¡
Peltier cooling insufficient.
¡
Inform the service.
¡
Replace the water traps after repair.
Error code: Error message
Minimum sample volume > cup volume
Cause
Remedy
¡
For sample supply via autosampler:
Selected sample volume too large.
Number of determinations too high.
¡
Check the following settings in the
method: Sample volume/purge volume.
¡
Adjust the number of determinations
(repeat measurements) to the vessel
volume.
7.2
Status errors
Status errors are displayed in red or yellow in the System state window.
Error indication
In 120 ml/min; Out < 110 ml/min
Cause
Remedy
¡
The union nut on the combustion tube is
not tightened correctly (after catalyst re-
placement).
¡
The carrier gas supply to the furnace
head or lock is not connected properly
(after catalyst replacement).
¡
The sealing rings on the combustion
tube are defective (heavily deformed) or
not attached (after catalyst replace-
ment).
¡
The FAST connector on the TIC conden-
sate container is leaky.
¡
The connections on the water trap sys-
tems are loose (after water trap replace-
ment or halogen trap maintenance).
¡
Check the screw connections for com-
pleteness and deformations. Tighten as
necessary.
¡
Check the carrier gas supply, in particu-
lar the FAST connector on the analyzer
wall and the screw connection on the
furnace head.
¡
Check all connection points on the water
traps. Replace the FAST connector as
necessary.
¡
The connection between the combustion
tube and the condensation coil or the
screw connections are leaky.
¡
Check the connection of the combustion
tube to the condensation coil, in particu-
lar the fork clamp position.
¡
The combustion tube is faulty (cracks,
fractures at the edge).
¡
TIC condensate container faulty (frac-
tures on the connections).
¡
Check the glass components. Replace as
necessary.
¡
Water traps clogged.
¡
Replace the water traps.
¡
The condensate pump hose is leaky.
¡
Check the condensate pump. Replace
the hose as necessary.
Error indication
In 120 ml/min; Out < 110 ml/min;
Out > 130 ml/min
Cause
Remedy
¡
The MFM (mass flow sensor) is faulty.
¡
Check the flow with an external mass
flow sensor if possible to confirm the er-
ror.
¡
Inform the service.
¡
The filling of the halogen trap is used
up.
¡
Check the halogen trap.