multi N/C pharma HT
Transport and storage
117
8.1.2
Preparing the AS vario autosampler for transport
NOTICE
Risk of device damage when transporting without transport locks
During transport without transport locks, the device can become damaged.
¡
Always apply a transport lock before transport.
Figure 66Securing the autosampler for transport
1 Transport lock
2 M3x12 screw
}
Turn the autosampler on its side and put it down safely.
}
Turn the autosampler arm clockwise up to the stop.
ü
The drives are in the correct position.
}
Slide the transport lock into the opening of the bottom plate up to the stop.
}
Fasten the transport lock with the screw and the supplied hexagon socket key.
}
Put the autosampler into its original packaging.
ü
The autosampler can now be safely transported.
8.1.3
Moving the device in the laboratory
CAUTION
Risk of injury during transport
Dropping the device poses a risk of injury and damage to the device.
¡
Proceed carefully when moving and transporting the device. Two persons are re-
quired to lift and carry the device.
¡
Grip the device firmly at the bottom with both hands and lift it simultaneously.