multi N/C pharma HT
Maintenance and care
79
6.2.2
Adjusting the AS vario autosampler
NOTICE
Risk of bending
The canula may bend during adjustment.
¡
Unscrew the screw connections on the canulas before adjustment.
}
Start the software. Wait for the device to initialize.
}
Open the window of the same name with the Instrument |Sampler Alignment menu
option.
}
Select needle in the list field in the Please select position needing adjustment group.
ü
The autosampler arm moves over the adjustment points on the sample tray.
Figure 51Adjusting the autosampler
}
Increase or decrease the z-axis values until the canulas are approx. 2 cm above the
adjustment points. Click the [needle adjust] button.
}
Align the canulas to the two adjustment points by carefully bending them.
Figure 52Adjustment points on the sample tray