Maintenance and care
multi N/C pharma HT
102
Figure 63Replacing the halogen trap
1 FAST connector to hose 2
2 FAST connector to hose 3
3 Clamp
4 Brass wool
5 Copper wool
}
Open the doors of the analyzer.
}
Remove the FAST connectors from the halogen trap and remove the U-tube from the
clamps.
}
Pull out the depleted copper wool or brass wool from the U-tube with tweezers or a
small hook.
}
Check the U-tube for cracks. Only reuse a fully intact U-tube.
}
If necessary, rinse the U-tube with ultrapure water and allow it to dry well.
}
Fill the U-tube with new copper wool and brass wool using tweezers or a small hook.
– Replace the complete contents of the U-tube. Do not pack the copper and brass
wool too tightly, but do not allow any larger empty spaces.
}
Cover the copper wool and the brass wool with cotton wool.
}
Carefully press the filled U-tube into the clamps again.
}
Reconnect the gas hoses with FAST connectors to the halogen trap:
– Hose 2 to the branch with copper wool (connection to the water trap)
– Hose 3 to the branch with brass wool (connection to the detector)
}
Check the system for leaks.
}
Close the doors of the analyzer again.