Installation and commissioning
multi N/C pharma HT
52
1
2
3
Figure 34Fingertight connection
1 Hose
2 Banjo bolt
3 Conical nipple
}
Connect the power supply unit to the mains network.
Checking and changing the
configuration
}
Switch on the analyzer. Start the multiWin program and initialize the analysis sys-
tem.
}
Check the device configuration with the Instrument | System-Information menu op-
tion in the multiWin set-up info window.
}
If necessary, change the configuration:
– Exit the multiWin program.
– Start the setup tool in the Windows interface with Start | Program files | multi-
Win | multiWin setup tool.
– Select the autosampler type in the Sampler list.
– Exit the multiWin set-up tool window with [Create].
– Start the multiWin program and initialize the analysis system.
– Open the Options window with menu option Configuration | Edit options. Go to
the Analyzer components tab.
– In the Sampler group, select the appropriate tablet and vessel size.
}
Exit the window with [OK].
Installing the canula flush
There is a suitable canula holder and a block with wash cups for each sample tray. The
tray, the canula holder and the block are marked with the maximum sample number,
e.g., 72.
Figure 35Canula flush on the AS vario ER model
1 Ultrapure water connection
2 Waste connection
3 Removable block with wash cups