multi N/C pharma HT
Operation
69
}
Use the calibration for the TOC/NPOC parameter for the lower concentration range.
A single point calibration is sufficient in the lower concentration range.
}
The software displays the sample feed type under Sample introduction. The indica-
tion is for information only and cannot be modified.
}
Select how to take the blank value of the prepared water of the standard solutions
into account in the Preparation blank group
– Measure selection field:
The content of the preparation water is measured separately immediately before
calibration. To do this, place a vessel with preparation water at the first position
on the autosampler.
With manual sample supply, the software sends a prompt to provide the prepara-
tion water.
– Enter selection field:
Enter the content of the preparation water as a value.
Specify the normalized preparation water blank value 1 ml. If the preparation wa-
ter blank value is not taken into account, a
"0"
is entered in the input field.
}
Complete the calibration table for each parameter in accordance with the standard
solutions provided.
}
In the Rep. column, the software automatically accepts the number of repeat mea-
surements set in the method. If outlier selection is enabled in the method, the maxi-
mum number of repeat measurements is entered.
If necessary, change the number of repeat measurements for each standard solution
manually.
}
If the calibration table is to be used again later: Save the calibration table with the
CalibrationTable |Save CalibrationTable or Save CalibrationTable as menu option.
Calibration tables are automatically assigned the *.kaltab file extension and are
stored under ...\Calibration\Tables.
}
Click the Start measurement button. Carry out the software instructions that follow.
ü
Depending on the selected method and sample supply type, further queries ap-
pear. With sample feed via autosampler, the Current sample data window opens.
Figure 47Current sample data
}
Activate the calibration standards in the Current sample data window. After this, exit
the window with [].
}
The Measurement window opens. Click the [Start F2] button.
ü
The calibration process starts.