Installation and commissioning
multi N/C pharma HT
46
1
2
3
Figure 27Fingertight connection
1 Hose
2 Banjo bolt
3 Conical nipple
}
Connect the power supply unit to the mains network.
Checking and changing the
configuration
}
Switch on the analyzer. Start the multiWin program and initialize the analysis sys-
tem.
}
Check the device configuration with the Instrument | System-Information menu op-
tion in the multiWin set-up info window.
}
If necessary, change the configuration:
– Exit the multiWin program.
– Start the setup tool in the Windows interface with Start | Program files | multi-
Win | multiWin setup tool.
– Select the autosampler type in the Sampler list.
– Exit the multiWin set-up tool window with [Create].
– Start the multiWin program and initialize the analysis system.
– Open the Options window with menu option Configuration | Edit options. Go to
the Analyzer components tab.
– In the Sampler group, select the appropriate tablet and vessel size.
}
Exit the window with [OK].
Adjusting the autosampler
During adjustment, the canulas are adjusted to the sample tray so that they are opti-
mally immersed into the sample vessels. Adjustment must be carried out during com-
missioning and after conversion, transport or storage.
}
Adjust the immersion depth of the canulas in the vessel (z-axis direction) in the soft-
ware.
– Open the window of the same name with the Instrument | Sampler Alignment
menu option.