Transport and storage
multi N/C pharma HT
118
Observe the following when moving the device within the laboratory:
¡
Insufficiently secured components pose a risk of injury!
Before moving the device, remove all loose parts and disconnect all connections
from the device.
¡
For safety reasons, two persons are required to transport the device, one person on
each side of the device.
¡
As the device does not have carrying handles, grip the device firmly with both hands
at the lower end. Lift the device simultaneously.
¡
Observe the guide values and adhere to the legally mandated limits for lifting and
carrying loads without auxiliary means.
¡
Observe the installation conditions at the new location.
8.2
Storage
NOTICE
Risk of device damage due to environmental conditions
Environmental influences and condensation can destroy individual components of the
device.
¡
Only store the device in air-conditioned rooms.
¡
Ensure that the atmosphere is free of dust and corrosive vapors.
If the device is not installed immediately after delivery or not required for longer periods,
it should be stored in its original packaging. A suitable desiccant should be added to the
equipment to prevent damage from moisture.
The requirements for the climatic conditions of the storage location can be found in the
specifications.