multi N/C pharma HT
Maintenance and care
95
}
Remove the top cover.
}
Remove the combustion tube. Slide the change-over valve to the right to
prevent it from interfering with removal.
}
Remove the condensation coil.
}
Pull the plug-in connector for the combustion furnace out of its socket.
}
Unscrew the knurled head screws the furnace is fastened to the device
bottom with.
}
Lift the furnace out of the analyzer.
}
Reattach the knurled head screws to the device bottom so they are not
lost.
See also
2
Removing the combustion tube [
2
Condensation coil maintenance [
6.8.2
Installing the combustion furnace
}
Open the left side wall of the analyzer. Remove the top cover.
}
Loosen the knurled head screw on the holder of the change-over valve.
Slide the change-over valve to the right to prevent it from interfering
with installation.
}
Unscrew the knurled head screws on the device bottom the furnace is
fastened with.
}
Insert the furnace in the middle. Align the front side of the furnace to be
parallel with the device wall.
}
Fasten the furnace with the knurled head screws. Tighten the knurled
head screws finger-tight.
}
Plug the plug-in connector for the combustion furnace into the socket at
the bottom right of the rear device wall.
}
Install the combustion furnace.
}
Install the condensation coil.
}
Push the sample intake hose and the purging hose through the top open-
ing. Attach the top cover. Connect the hoses to the autosampler.
}
Close the side wall.
– Connect the protective grounding to the left side wall.
– Slightly tighten the screws first on the bottom side and then on the
top side. Tighten the screws in turns.
}
Open the gas supply. Connect the power plug with the socket and switch
on the analyzer via the main switch.
}
Check the system for leaks.
ü
The analyzer is ready for operation again.