Installation and commissioning
multi N/C pharma HT
44
4.3.1.1 Installing and commissioning the sampler
CAUTION
Risk of injury from moving parts
There is a risk of injury in the movement range of the sampler arm. For example, hands
or fingers might be crushed.
¡
Maintain a safety distance from the sampler during operation.
NOTICE
Risk of device damage
If the sampler arm is obstructed during operation, the drives can be destroyed.
¡
Do not touch the sampler arm during operation.
¡
Only carry out manual adjustment when the device is switched off.
}
Switch off the analyzer before installing the autosampler.
}
Plug the grounding conductor into the connection on the rear of the analyzer. Con-
nect the grounding conductor to the connection on the bottom of the sampler.
}
Plug the cable on the low voltage side of the external power supply unit into the con-
nection on the bottom of the sampler. Do not connect the power supply unit to the
mains power supply yet.
}
Connect the autosampler to the analyzer with the interface cable (interface on the
bottom of the sampler and "sampler" interface on the rear of the analyzer).
1
2
3
Figure 25Connections on the bottom of the autosampler
1 Connection for equipotential bonding
cable (grounding cable)
2 Power cable connection
3 Analyzer interface