multi N/C pharma HT
Installation and commissioning
53
}
Place the suitable block with wash cups on the autosampler.
– For simpler installation, wet the o-ring on the bottom of the block with water.
– Fasten the block to the autosampler with the two hexagon socket screws.
}
Screw the ultrapure water connection into connection (1) and place the hose end in
the ultrapure water bottle.
}
Insert the waste hose in connection (2). Place the hose end in the waste container.
NOTICE! Route the outlet tube at a constant downward incline. If necessary,
shorten the hose. The hose must not dip into the liquid.
}
Adjust the autosampler before the first start.
Activating the canula flush for
measurements
Figure 36Activating the canula flush in the method
}
Create a new method.
}
Activate the Reverse rinse option under Variable method criteria.
}
Define the number of flushing cycles by entering a number ≥1 in the table. One
flushing process is usually sufficient.
See also
2