multi N/C pharma HT
Maintenance and care
81
Figure 53Installing canulas (2 canulas for parallel purging here)
}
Install the hold-down clamp and the sample intake canulas in the canula holder. Un-
screw the retaining screws of the canulas before adjustment. Clamp the canulas into
the holder so that the canula tip does not immerse in the sample vessel.
}
Open the window of the same name with the Instrument |Sampler Alignment menu
option.
}
Adjust the immersion depth of the sample intake canula into the wash cup and into a
sample vessel in position 1 of the sample tray.
}
To do this, select the Rinse position or the Position 1 entry in the Please select posi-
tion needing adjustment group.
Figure 54Adjusting the autosampler
}
Increase or decrease the x-, y- and z-axis values to align the flushing position or po-
sition 1: