multi N/C pharma HT
Installation and commissioning
49
Sample tray
Max. number of samples
Sample vessel
AS vario
AS vario ER
20
100 ml
Yes
No
52
100 ml
Yes
No
72
40 ml + 50 ml
Yes
Yes
100
20 ml
Yes
Yes
146
12 ml
Yes
Yes
Technical data
Operating voltage
24 V DC via external power supply
Power consumption
50 VA
Grid voltage of external power supply
100 to 240 V, 50/60 Hz (autosensing)
Dimensions (W x H x D)
350 x 470 x 400 mm
The autosampler is placed next to the analyzer. It can be equipped with 2 canulas.
Figure 30Layout of the AS vario autosampler
1 Connection hose to the analyzer (purg-
ing hose for NPOC measurement)
2 Connection hose to the analyzer (sam-
ple intake hose)
3 Canula holder
4 Autosampler arm
5 Sample vessel
6 Sample tray
7 Sleeve
8 Canula
The AS vario ER model is particularly suited for analyzing liquid samples with high solid
particle content. The model is equipped with an additional canula flush that flushes the
canula(s) with ultrapure water from the outside. When commissioning the autosampler,
the ultrapure water supply for the canula flush must be installed additionally. It can be