Figure 3-11: Installation of Verification Pin and Inclination Requirements for Flange
and Still-pipe
)
,
0
+
*
&
'
(
.
-
%
$
A. 1000 < L < 2500 mm (39 < L < 98 in.).
Recommended: 1200 mm (47 in.)
G. The Verification Pin is directed towards the
bolt hole at the pipe flange marking.
B. Min. 200 mm (8 in.) from Verification pin to
product
H. Bolt hole
C. Hole for Verification Pin; Ø 20 mm.
I.
Align Verification Pin and bolt hole within
1°.
D. Holes for density equalization; Ø 20 mm
(3/4 in.)
J.
Maximum 1°
E.
500 mm (20 in.)
K. Maximum 0.5°
F.
Mark on still-pipe flange