![Emerson Rosemount 5900S Series Скачать руководство пользователя страница 290](http://html.mh-extra.com/html/emerson/rosemount-5900s-series/rosemount-5900s-series_reference-manual_101196290.webp)
Table C-18: Advanced Configuration Transducer Block Parameters
(continued)
Index
Number
Parameter
Description
6
BLOCK_ERR
This parameter reflects the error status associated
with the hardware or software components
associated with a block. It is a bit string, so that
multiple errors may be shown.
7
UPDATE_EVT
This alert is generated by any change to the static
data
8
BLOCK_ALM
The block alarm is used for all configuration,
hardware, connection failure or system problems
in the block. The cause of the alert is entered in
the subcode field. The first alert to become active
will set the Active status in the Status parameter.
As soon as the Unreported status is cleared by the
alert reporting task, another block alert may be
reported without clearing the Active status, if the
subcode has changed.
9
TRANSDUCER_DIRECTORY
Directory that specifies the number and starting
indices of the transducers in the transducer block.
10
TRANSDUCER_TYPE
Identifies the transducer.
11
TRANSDUCER_TYPE_VER
12
XD_ERROR
A transducer block alarm sub code.
13
COLLECTION_DIRECTORY
A directory that specifies the number, starting
indices, and DD Item ID’s of the data collections in
each transducer within a transducer block.
14
AUTO_CONF_MEAS_FUNC
Check box to enable manual settings of affected
parameters
15
USED_EXTRA_ECHO_MIN_ULLAG
E
Parameters and functions for Empty Tank
Handling.
See
16
USED_EXTRA_ECHO_MAX_ULLAG
E
17
USED_EXTRA_ECHO_MIN_AMPL
18
EXTRA_ECHO_MIN_ULLAGE
19
EXTRA_ECHO_MAX_ULLAGE
20
EXTRA_ECHO_MIN_AMPL
21
USED_EMPTY_TANK_DETECTION_
AREA
22
EMPTY_TANK_DETECTION_AREA
23
USED_ECHO_TIMEOUT
Parameters and functions for Echo Tracking.
See
for more information.
24
USED_CLOSE_DIST
25
USED_SLOW_SEARCH_SPEED
26
USED_FFT_MATCH_THRESH
F
OUNDATION
™
Fieldbus Block Information
Reference Manual
September 2019
00809-0100-5900
290
Reference Manual
Содержание Rosemount 5900S Series
Страница 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Страница 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Страница 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Страница 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Страница 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Страница 222: ...Service and troubleshooting Reference Manual September 2019 00809 0100 5900 222 Reference Manual ...
Страница 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...